 MA8500  Neonatal Monitor

Test Result

SunTech Advantage NIBP

Averaged deviation of Systolic pressure

-2.56 mmHg

Standard deviation of Systolic pressure

5.71 mmHg

Averaged deviation of diastolic pressure

-4.51 mmHg

Standard deviation of diastolic pressure

5.57 mmHg

 

Test Methods: Participants were 30 newborn baby who were measured 180 groups data 

                         for systolic (SBP) and diastolic (DBP) blood pressure锛 each NIBP measurement reactivity 

                         to relevant IBP measure form indwelling catheter in umbilical artery or femoral artery , 

              Averaged deviation should be less than ±5mmHg and the standard deviation should be less than 8mmHg.

                         This clinical test methods follow the standard procedure of AAMI SP10:2002/A1:2003

 

 
 
 
 
 
 