is leading exporter and supply chain
service provider on medical equipment.
MY-C005 12.1inch patient monitor

Product Type: patient monitor
Place of origin: China (Mainland) Guangdong
Model No: MY-C005
Price Terms: FOB Guangzhou
Payment Terms: T/T.L/C.Western Union.MoneyGram
Package: standard export packing
Minimum Order: 1 set/sets
Delivery Time: 15 days after receive payment
Brand Name: MAYA

 

Features:

12.1'' screen with true color, wide viewing angle, high brightness LCD display.

HR/PR, NIBP, RR, ECG, SpO2, TEMP

Simple and friendly operating display interface.

Rechargeable lithium battery provides convenience for patients' moving.

Playback and browse function for long term waveform and monitor data record.

Optional printing output function, alarm triggers printing.

Auto double alarm with audible and visible signals

Anti-defibrillation, anti-interference from high frequency electric knife

A full-synchronistic lead multi-channel ECG display...

This medical equipment can be used in ICU, CCU and operation room.

Suitable for adult,pediatrics and neonate patients. 

Standard configuration:

HR/PR, NIBP, RR, ECG, SpO2, TEMP, Rechargeable lithium battery

Optional:

Dual TEMP

Dual IBP

ETCO2

Thermal recorder

Wall mount

Trolley

Specification:

ECG

Lead configuration

Standard 3-lead or 5-lead 

3-leadRA,LA,LL,Lead method:I,II,III

5-leadRA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V

Increase ´250,´500,´1000,´2000

HR

Range

   Adult                   15 ~ 300bpm (beat/minute)

   Neonatal baby/children       15 ~ 350 bpm(beat/minute)

Precision±1% or ±1bpm,the larger prevails

Resolution   1 bpm(beat/minute)

Sensitivity 

> 200 uV (Peak-to-peak value)

Input Impedance

> 5 (megohm)

Bandwidth

Diagnostic mode0.05~130Hz

Monitoring Mode    0.5~40Hz

Operation mode1~20Hz

Common Mode rejection Ratio

Diagnostic mode> 90 dB

Monitoring Mode    > 100 dB

Operation mode> 100 dB

Pole Polarization Vvoltage Range ±300mV

Pacing Pulse Test

Test pacing pulse in accordance with the following conditions:

Amplitude:±2 mV ~ ±700mV

Width:    0.1ms ~ 2ms

Risetime:      10us~100µs

Pacing pulse inhibition

Amplitude:±2 mV ~ ±700mV

Width:   0.1ms ~ 2ms

Risetime:     10us ~ 100µs

Baseline Recovering Time

After defibrillation< 3 seconds 

Signal Range

±8 mV (Peak-to-peak value)

Calibrating Signal

1mV(Peak-to-peak value),precision±5%

ST Segment Measuring Volume

Measuring range: -2.0mV ~ +2.0mV

No definition for other ranges.

RESP 

Measuring Method

RA-LL impedance

RESP Impedance Measuring Range

0.3~

Base Impedance Range

200~4000Ω

Bandwidth

0.1~2.5Hz

RESP Rate

Range

Adult          0~120BrPM

Children and neonatal baby 0~150 BrPM

Resolution    1 BrPM

Precision    ±2 BrPM

Asphyxia Alarm

10~40 seconds

NIBP 

Measuring Method

Pulse wave oscillometry 

Work Mode

Manual/Automatic/STAT

Measuring Interval of Automatic Measuring Mode

1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)

Measuring Time of STAT Mode

5 minutes

PR range

40 – 240 bpm

Overvoltage protection

Adult mode        300 mmHg

Children mode        240 mmHg

Neonatal baby mode150 mmHg

MY-C001 Color Handheld Multi-Parameter Touch screen patient MY-C011 Medical patient monitor 12.1 Fetal Patient Monitor
Recommended Products

Contact Us

  • Room 607, No.1 Yingbin Building,Shangjiao,Dashi Town,Panyu Area,Guangzhou,China
  • Tel:86-20--39160381/ 39154381
    Fax:86-20-84597079
    Mobile:86-18925044116
  • webmaster@medicalequipment168.com
    nancy@medicalequipment168.com
  • nancy@bultrasound.com